Dermlite DL100 - Directions for Use
Intended Use
The DermLite DL100 is designed for medical professionals to perform non-invasive visual skin examinations using cross-polarized light.
CAUTION: Do not look directly into the LED light. Patients must close their eyes during examinations.
CAUTION: Not for use in fire- or explosion-prone environments.
WARNING: May expose users to substances known to cause cancer or reproductive harm (e.g., methylene chloride, hexavalent chromium).
Device Overview
The DL100 is a handheld, battery-powered dermatoscope that uses eight 3 mm white LEDs and a 10x magnification lens with cross-polarization.
Operation
- Position the DermLite DL100 approximately 12 mm (0.5”) from the skin.
- Use your fingers or the included spacer to maintain optimal distance.
- Press the power button on the side to activate the LEDs; hold the button to keep light on.
- Move the device closer or farther from the lesion to focus the image.
Battery Replacement
- Squeeze the battery cap to remove it along with the battery.
- Insert a new 2CR5 lithium battery with contacts facing forward.
- Reattach the battery cap securely.
Battery Details
Use only 2CR5 lithium batteries from DermLite or a trusted retailer. Battery is warranted for 1 year.
Troubleshooting
For support, visit www.dermlite.com/service or contact your DermLite dealer.
Maintenance
CAUTION: Modification of this device is not permitted.
Clean the exterior with 70% isopropyl alcohol. Avoid submerging or autoclaving the device. Clean the lens with standard optical tools.
Warranty
5 years for parts and labor. Battery warranty is 1 year.
Disposal
Dispose of lithium-ion battery and electronic components separately per local regulations.
Package Contents
- DermLite DL100 unit
- Eight LEDs with cross-polarization
- 10x magnification lens
- 2CR5 lithium battery
- Removable spacer